Step 1 Follow provided
directions for sampling and cracking the ampoule. Be sure to shake
ampoule well to get all components of test into solution.
Step 2 Remove contents
(without rod) into a test tube. Contents can be removed aseptically
from opening in top with a needle and syringe. Or the outside of the
ampoule can be cleaned with alcohol and then the top cracked open
with a sterilized tool.
Step 3 Make a small
ink mark at the top of the test tube.
Step 4 Set
spectrophotometer to 465nm.
Step 5 Insert test
tube into spectrophotometer with ink mark toward front and take a
"time zero" reading of % Transmittance. Record this result
along with the time.
Step 6 Incubate at
35-37C.
Step 7 Take a spectrophotometric
reading every half hour making sure ink mark is toward the front.
Record results. When there is a 10% reduction in %T the test is positive.
(ie-75%T and 5 hours 65%T)
Step 8 Add 1/2 hour
to the time in Step 7 and use visual results timetable to determine
cfu/ml in the sample.
**NOTE**
This method works for counts of 10 to 1,000,000 cfu/ml. For counts
higher than 1,000,000 cfu/ml it is best to use the visual test procedure
or perform serial dilutions of the sample. Serial dilutions lower
the bacterial count in the sample tested and enable you to use the
spectrophotometric procedure. To obtain a final result simply multiply
by the dilution factor. (example- a 1/100 dilution yields a 10% reduction
in %Transmittance after 5 hours; using the visual results timetable
for 5.5 hours the count in sample tested is 100,000 cfu/ml; multiply
this result by 100 to get a count of 10,000,000 cfu/ml in the actual
sample) **Please note that floating particulates in test sample can
interfere with spectrophotometric readings. The sample can be filtered
before beginning the test to remove particulates. Only filter sample
if the floating matter is not necessary for your specific test. (ie-containing
a biologic you are testing for) If the sample can not be filtered
please use the visual method.
|