|
|
|
The following questions are typical
of what might be encountered by first time field users. The questions
have been organized by test function to assist sales and customer
service personnel with the task of providing answers. Most of these
questions and answers apply directly to the newer Coliform (CF KOUNT
Assayer) and E. coli (EC KOUNT Assayer).
-
HOW LARGE OF A SAMPLE DO I NEED?
It is recommended that the provided sample cup be filled to
the top (approximately 30 mls) with sample water. The KKA ampoule
requires 8 mls to fill. Because the ampoule is under vacuum
and draws a sample quickly, it is necessary to have more than
the required 8 mls in order to prevent the sucking of air which
will result in an under filled KKA.
-
DOES
IT MATTER HOW I TAKE THE SAMPLE? Yes! bacteria can cluster
in water. This is especially true of waters which have no movement.
If the water to be sampled is agitated the distribution of bacteria
can be expected to be even. Accordingly, it is a good idea to run
a little water out of the reservoir to be sampled before drawing
a sample.
-
HOW
IMPORTANT IS IT THAT THE SAMPLE CUP BE KEPT CLEAN AND WHAT IS MEANT
BY "CLEAN"? It is important to keep the sample
cup as clean as possible. The residence time (15 minutes) of the
sample water in the sample cup is usually long enough to pick up
side wall residues be they biologic or chemical. A simple rinsing
of the sample cup with the water to be sampled (2-3 times) will
normally eliminate sample cup interferences. When not in use, keep
the sample cup in a dust free and dry area. For high sensitivity
(l-10 cfu/mL) laboratory measurements, the sample cup maybe sterilized
by rinsing the cup with tap or deionized water and microwaving for
20-30 seconds the sample cup with the final rinse water in it.
-
IF
I AM RUNNING A CONTROL AMPOULE WITH KKA HOW MUCH SAMPLE IS REQUIRED?
The control ampoule uses the same amount of sample water as the
KKA (8 mls). Therefore, to run a control and KKA together 16 mls
of sample water would be required. A completely filled sample cup
will contain enough sample water for both - 1 control and 1 KKA
ampoule.
-
WHY
DO I NEED TO LET THE WATER SAMPLE SIT IN THE SAMPLING CUP FOR 15
MINUTES? Most of the time it is not necessary but letting
the water sit is a passive form of sample conditioning. If the water
contains low levels of halogens (chlorine, bromine) they can dissipate.
Suspected non-biologic matter can settle out. Extremely high or
low temperature samples will moderate.
-
WHAT
IF MY WATER SAMPLE IS NOT CLEAR? For turbid or colored water
a Control Ampoule should be used at the same time as the KKA ampoule.
Both Control and KKA should be filled with the same sample water.
The Control ampoule will assist in interpreting end of test colors.
Samples (wastewater, etc) can be diluted to lessen the effect of
color of turbidity.
-
WHAT
IS THE DIFFERENCE BETWEEN A SAMPLE CUP AND A SNAPPING CUP?
Nothing! they are the same item. Each kit comes with a sample/snapping
cup.
|
-
DOES IT TAKE A LOT OF PRESSURE TO
SNAP THE KKA TIP? No, KKA ampoules
will snap easily. The tip of the ampoule is scored to facilitate
the snapping of the tip. By rotating the ampoule between your
thumb and forefinger while pulling the ampoule towards the opposite
wall of the sample cup, the scored snapping point will be reached
causing an easy break. After snapping 3-4 ampoules, you will get
the "FEEL" for snapping.
-
DO
I HAVE TO USE THE PROVIDED AMPOULE SNAPPING CUP? No, the
snapping/sample cup is provided to facilitate ampoule snapping.
Each box comes with a new cup. Any clean container, large enough
to hold 30 mls and wide enough to permit snapping using the side
wall as a prying surface will work.
-
WHAT SHOULD I DO
IF KKA DOES NOT FILL COMPLETELY OR AT ALL?
If KKA does not fill completely, air was sucked in at the time
of filling. This situation occurs when too little sample water
is in the snapping cup or the ampoule is withdrawn too quickly.
If the ampoule is filled to within 0.5 inches of the top of the
ampoule shoulder continue the test. If the fill is less than 0.5
inches of the top of the shoulder abandon the test and run another
ampoule making sure to have enough sample water in the snapping
cup so that air is not sucked into the ampoule or hold the ampoule
in the water until it fills.
-
IF
A CONTROL AMPOULE IS BEING USED WHEN DO I FILL IT? Control
ampoules should be filled with the same sample water immediately
before or after the KKA.
-
HOW LONG DO I KEEP
THE KKA AMPOULE IN THE SAMPLING CUP?
As soon as the KKA ampoule has filled it can be removed. Filling
takes 2-3 seconds.
-
WHAT
DO I DO AFTER THE KKA AMPOULE IS FILLED? Rock the KKA ampoule
in your hand until the powder has dissolved and go to incubation
portion of the instructions.
-
WHAT DO I DO IF THE
POWDER IS NOT DISSOLVING? If the
sample water temperature is significantly below the incubation
temperature it will take longer to dissolve all of the powder.
Placing the KKA ampoule in a warm environment (same as an incubator)
for a brief period and than reshaking to mix will solve the problem.
-
DO
I NEED AN INCUBATOR? No, KKA was designed to be user friendly
in the FIELD or LAB. Placing the KKA in any type of environment
(shirt pocket, heater room, etc.) which is a constant 35 C / 95
F for the time period the test is running is all that is required.
-
IS IT NECESSARY TO
INCUBATE KKA AT EXACTLY 35 C / 95 F?
No, incubation temperature may vary 2-3 degrees either side of the
specified temperature.
-
WHAT
DO I DO IF THE INCUBATION IS INTERRUPTED BY POWER FAILURE OR OTHER
CAUSES? Incubation can be interrupted for up to 48 hours.
During the interruption KKA should be placed in a cool environment
(40-50 F); when you are ready to restart the test place KKA back
at incubation temperature and follow the stop/restart instructions.
-
WHAT
WILL HAPPEN IF KKA IS INCUBATED AT LOWER THAN SPECIFIED TEMPERATURES?
KKA will develop to end of test but at a slower rate. How much slower
is directly proportional to the temperature differential from the
specified incubation temperature. It is possible to recalibrate
for lower temperatures.
-
WHAT
WILL HAPPEN IF KKA IS INCUBATED AT HIGHER THAN SPECIFIED TEMPERATURES?
KKA will develop to end of test at a quicker rate than the development
times specified. Using higher temperatures should be practiced by
only experienced personnel since too much heat can kill some microorganisms.
-
CAN
I STOP AND RESTART INCUBATION, AND IF SO FOR HOW LONG? Yes,
placing the ampoule in a refrigerator will stop the reaction. When
removing KKA from the refrigerator to restart the test, add 1/2
hour to the elapsed development time to allow the sample water to
return to proper incubation temperature. KKA can be refrigerated
up to 24 hours without any problems.
-
HOW
LONG DOES IT TAKE FOR KKA TO PASS THROUGH A COLOR PHASE?
It takes KKA about 30 minutes to complete a color change once the
orange color has been reached. Extremely high counts would make
the time in each color shorter.
-
WHAT
HAPPENS IF I DO NOT VIEW THE KKA THROUGH THE BOTTOM? Not
viewing the ampoule from the bottom will cause the operator to miss
or not see the required color changes at the correct time. The result
will be significantly low, erroneous readings.
-
WHAT
DO I DO IF I MISS THE ORANGE COLOR WHEN USING THE PROFILE METHOD?
It is really not a problem; because color phases take a predicable
amount of time (30 min.), the time to orange can be calculated by
deducting 30 min. for each color phase not seen. The instructions
for this method provide additional information.
-
HOW
DO I USE A CONTROL AMPOULE? The control ampoule is held right
next to the KKA ampoule and both are viewed together through the
bottom. The control ampoule provides the comparison or starting
point for end of test color determination. For colored or turbid
sample waters, a control ampoule is very useful.
-
ARE
THERE ANY CHEMICALS WHICH INTERFERE WITH THE KKA REACTION?
Yes, chemicals which are strong reducers will interfere with the
test. Strong reducing agents can give an early red color development
(1-5 min.). Sample pretreatment will depend on the compound. A general
nuetralizing agent is available as a seperate purchase item. If
your desire is to have a "presence" count (as opposed
to an "activity" count) of the bacteria and halogens are
being used, a simple addition of 1 or 2 drops of sodium thiosulfate
(supplied with kit) will neutralize halogens.
-
DO BIOCIDES INTERFERE
WITH THE KKA REACTION? Generally
not when used in concentrations of 100 ppm or less. There are some
biocides like Glutaraldehyde, which virtually eliminate most microbes
except Pseudo. species. Generally, such situations require an additional
2 hours of KKA development time (see KKA-W Time chart).
-
WHAT
DOES IT MEAN IF I SEE ORANGE OR PINK IN 5 MINUTES OR LESS?
This usually means the presence of a strong reducing chemical, especially
if the color goes to an immediate light pink (or darker) which does
not develop further. Or, if the color keeps developing to red and
the test is being done outside, the ultra voilet light (sunlight)
is the problem. The correction for outside testing in sunlight is
to use the light blocking black sleeve provided with each test kit.
-
WHAT
DO I DO IF KKA RESULTS ARE SLOWER OR LOWER THAN EXPECTED?
Factors which can contribute to slower development times (low results)
are low incubation temperatures, certain biocides or a predominance
of one bacterium species.
-
WHAT
DOES IT MEAN IF I GET NO COLOR CHANGE? The count is exceptionally
low or there is a strong oxidizing chemical present. Always check
to see if a stir bar is present. If not, a Control Ampoule was used
and it is not supposed to change color.
-
IF
I GET UNEXPECTED RESULTS SHOULD I SAVE THE KKA AND HOW? One
of the distinguishing features of KKA is to be able to retain a
sample which has given unusual results. It is best to refrigerate
or ice the KKA sample until it can be analyzed by sophisticated
laboratory techniques for chemical and microbial content.
-
AFTER
I HAVE FINISHED USING THE KKA HOW DO I DISPOSE OF IT? There
are several methods for proper disposal of used KKA ampoules. Basically,
the same methods used for agar paddles would apply. In the field,
one approach would be to clip the open ampoule end (with a device
like a nail clipper) and dump the contents back into the same body
of water from which the sample was taken, or inject chlorine or
bleach into the ampoule to kill the organisms and dispose of the
ampoule contents into the same water body from which the sample
was taken. Retain the empty glass ampoule for autoclaving and destruction
at a convenient time.
-
IF
KKA IS DROPPED AND BROKEN WHAT PRECAUTIONS SHOULD BE TAKEN IN CLEANING
UP THE MESS? Spill and cleanup information is contained on
the product MSDS.
-
ARE
KKA REAGENTS PERSONALLY DANGEROUS MATERIAL? No!
-
IS
KKA CONSIDERED BY DOT / OSHA A HAZARDOUS PRODUCT? No!
-
WHAT
SAFETY EQUIPMENT IS REQUIRED TO USE KKA? None!
-
HOW IMPORTANT IS IT
TO USE THE PROVIDED PLASTIC HOSES?
Very! the plastic hose is applied over the broken end to prevent
cutting of skin or other surfaces. In addition, a carry/incubation
tube is also available.
-
WHERE
WOULD I FIND ALL APPROPRIATE EMERGENCY MEDICAL INFORMATION FOR KKA?
The MSDS contains all appropriate emergency information.
-
WHAT
ARE CONTROL AMPOULES AND WHEN DO I NEED THEM? Control ampoules
are those without the stir bar. They contain no indicator and will
not develop a color change past the yellow start color point. They
are used for sample waters which contain color or turbidity to help
the operator see a color change in the KKA ampoule. They act as
a constant starting point so that changes can be easily seen in
the KKA.
-
HOW
DO I OBTAIN CONTROL AMPOULES? The may be ordered in a box
of 8.
-
WHEN
DO I FILL A CONTROL AMPOULE? Control ampoules should be filled
at the same time as the KKA ampoule and with the same sample water.
-
HOW
LONG WILL A CONTROL AMPOULE LAST? Control ampoules will last
30 days or more for most sample waters especially if it is kept
cool (refrigerated). If the sample water remains constant for color
and turbidity, one control ampoule can be used repeatedly for many
KKA ampoules.
-
HOW
DO I DISTINGUISH A CONTROL AMPOULE FROM A KKA AMPOULE? Control
ampoules have no stir bar, unlike the KKA which does have an internal
stir bar.
-
CAN
KKA BE USED FOR WATER OTHER THAN COOLING TOWER WATERS? Yes,
KKA can be used for virtually any viable microbe count in an aqueous
solution.
-
WHAT
ORGANISMS WILL BE DETECTED BY KKA? KKA measures all viable
aerobic microbes. If they are growing in the solution they will
grow in the KKA. Times of development can vary for each application
and/or microbe type.
-
CAN KKA BE USED FOR
SINGLE SPECIES ANALYSIS? Yes, by
using various sample pretreatments and selected incubation temperatures
which isolate certain bacteria species, KKA can perform single species
analysis. Such programs need to be specifically engineered by application.
-
HOW
DO I CALIBRATE KKA FOR NON COOLING TOWER WATERS? A simple
method is to run a series of KKA ampoules parallel to the existing
technique. Record the times for development of the end of test color
and correlate these times to the values the older, slower method
records. Do this for several value levels and you will have made
a viable calibration.
-
WHY
ARE THERE 2 METHODS OF OPERATION FOR KKA? For most Cooling
Tower applications, a specific control point is known or desired.
For these applications, the Limit Control Method is most desirable
because an operator need only look at the KKA once. For applications
where there is no known or expected count, the Profile Method is
required.
-
CAN
A FILTER PHOTOMETER OR SPECTROPHOTOMETER BE USED TO MEASURE COLOR
DEVELOPMENT? Yes, if you desire to use non-visual methods
for end point color determination select an instrument which can
accommodate the KKA. It will need to be calibrated.
-
WHAT
ARE THE BEST AND WORST STORAGE CONDITIONS FOR KKA? KKA is
best stored in the box in an upright position. Ideal temperatures
are room temperature or below. The worst environment is to store
KKA out of the box, on its side, in intense UV light with temperatures
over 120 F.
-
BECAUSE
KKA IS MADE OF GLASS ARE THERE ANY SPECIAL HANDLING PROCEDURES REQUIRED?
When stored in the KKA box it is a rugged, durable product. Outside
of the box, care should be exercised to prevent the tip from receiving
lateral pressure which would cause premature breakage of the tip.
-
HOW
DO I OBTAIN REPLACEMENT PARTS; i.e., SNAPPING CUP, COLOR CHARTS?
Each KKA kit comes complete with all necessary parts.
|